The AQ5000Pro handheld IAQ monitor
is designed for building walk through surveys. It's portable design allows
you to quickly respond to your customers indoor air quality complaints. Thus it is more convenient than other internal air quality monitors.
To
provide an accurate indoor air quality picture, the AQ5000Pro IAQ meter records
carbon dioxide, relative humidity, temperature, and dew point. Available
options include support for additional gases and air velocity. A linear analog input channel interfaces with external
devices including VOC monitors, respirable dust monitors, light meters,
formaldehyde monitors and more.
After you complete your survey and have captured your indoor air meter data, you can print a
hardcopy for review or you can download the data into QuestSuite Professional. QuestSuite Professional software is simple to use and allows you to analyze, chart,
report and archive your data.
For pricing and availability information regarding the
purchase, service or calibration of the AQ5000Pro IAQ monitor, please email Customer Service.
For answers to your questions regarding installation, setup, operation,
maintenance or trouble-shooting of the AQ5000Pro IAQ monitor, please email Technical Support.
Brochures Available:
English Version (pdf) 1.9M
Spanish Version (pdf) 750K

AQ5000Pro
IAQ Monitor Technical Specifications
Measured Parameters:
Carbon Dioxide:
- Detector NDIR (non-dispersive infrared)
- Range User Selectable
- 0 to 5,000 and 0 to 20,000 ppm
- Accuracy 0 to 5,000 ppm: +/-3% of reading +/-50 ppm
- 0 to 20,000 ppm: +/-3% of reading +/-300 ppm
Temperature Influence Additional +/-0.2% of reading
+/-2 ppm for each °C
- Resolution 1 ppm (10 ppm above 10,000 ppm)
- Response Time 22 seconds for 63% of final value; 70 seconds for
10-90% of 4000 ppm change
Relative Humidity:
- Detector Capacitive
- Range 0-100% non-condensing
- Accuracy +/-3% @ 25°C
- Resolution 0.1%
- Response Time 50 seconds with min air flow of 1 mi/hr
- (.45 meters/sec) through wand
Temperature:
- Detector Thermistor
- Range 0 to 60°C (+32 to +140°F)
- Accuracy +/-5°C (+/-0.9 °F) requires min air flow of 1 mi/hr
(.45 meters/sec) through wand
- Resolution 0.1 oC
Dew Point:
- Range -65 °C/+60 °C
- Resolution 0.1 °C
Gas:
- Detector O2, CO, H2S, SO2, CL2, NO2, NO, HCN & NH3
- Resolution 0.1, 0.5 or 1 ppm depending on sensor in use
Air Velocity:
- Detector Hot-Wire
- Range 0-20 m/s
- Accuracy +/- 5%
- Resolution 0.1 m/s
Linear:
- Range +/-1 V
- Accuracy +/-2% of full scale
- Resolution 0.001 Vdc
Data Storage:
Display:
- 20 character x 4 lines backlit LCD
Real Time Clock:
Storage Rate:
- 1 sec, 1 min, 5 min, 10 min or 15 min sample or averages
Statistics Saved:
- Up to 5 data values per sampling interval (CO2, temperature,
humidity, dew point, gas and linear and air velocity input)
Communications:
- Format RS-232, 1 start, 8 data, 1 stop bit, no parity
- Baud Rate 1200 to 38400
Printer Output:
Summary Report includes:
- Separate reports for each logging session indicating
minimum, maximum & time of occurrence, elapsed time &
average per session.
- Gas readings include:
TWA & STEL
Time History Report Includes:
All recorded data values with date & time stamp
Power:
- Internal Power:
"AA" Alkaline Batteries - >15 hrs @ 25°C (77°F)
- Rechargeable NiMH Battery Pack: 7 hrs @ 25°C (77°F)
(Data Retention 30 days in the event of low battery)
- External Power:
+6.0 Vdc output AC power adapter, 120 VAC or 240v
Environments:
- Operating Temperature: 0 to +50°C (+32 to +122°F)
- Storage Temperature: -20 to +60°C (-4 to +140°F)
Size:
- Enclosure Size 3.6 x 7.1 x 1.3" (9.1 x 18.0 x 3.3 cm)
- Wand Size 11.8" (30cm) long, 1" (2.54 cm) diameter
Weight:
 |
|